戊胺CH3(CH2)3CH2NH2气体检测管105SD,戊胺CH3(CH2)3CH2NH2气体检测管105SD PENTYL AMINE 2-22ppm,日本光明理化工业株式会社从1947年开始,在气体分析领域开拓了六十余年 , 形成了驰名于世的 ” 北川 ” 式气体检测管。气体快速分析法 ” 北川 ” 的历史就是日本检测管的历史 。目前是世界上*大的检测管生产厂家,也是日本**家通过美国SEI的认证,欧盟的认证的检测管厂家。公司不仅有ISO9001:2000的认证,还获得了环境管理的ISO14001的认证。北川式手动气体采集器AP-20 备有采气完成指示计 , 可以清楚的判明采气是否完成 . • 有专门截留玻璃的截留槽 , 以防止检测管割尖时玻璃片飞溅 . • 有采气容积为 50 毫升和 100 毫升的两段式制动卡 . • 采气器极为小巧 , 并有便于携带的皮具包 。
Tube No.105SD
1. PERFORMANCE
1) Measuring range :1-20 ppm 0.2-1 ppm
Number of pump strokes 1(100ml) 5(500ml)
2) Sampling time :1 minute/1 pump stroke
3) Detectable limit :0.1 ppm(100ml)
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :Pale purple→Pale Yellow
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
By reacting with Phosphoric acid, PH indicator is discoloured.
2NH3+H3PO4→(NH4)2PO4
4. CALIBRATION OF THE TUBE
PERMEATION TUBE METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
20
18
16
14
12
10
8
6
4
2 1
100ml
ppm
GAS To pump
DEHUMIDIFIER
(White)
105SD AMMONIA
Substance Interference Coexistence
Amines Similar stain is produced. Higher readings are given.
(NOTE)
When the concentration is below 1 ppm, 2 to 5 pump strokes can be used to detemine the lower concentration.
Following formula is available for actual concentration.
Actual concentration=Reading value× 1
Number of strokes
1. PERFORMANCE
1) Measuring range :1-20 ppm 0.2-1 ppm
Number of pump strokes 1(100ml) 5(500ml)
2) Sampling time :1 minute/1 pump stroke
3) Detectable limit :0.1 ppm(100ml)
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :Pale purple→Pale Yellow
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
By reacting with Phosphoric acid, PH indicator is discoloured.
2NH3+H3PO4→(NH4)2PO4
4. CALIBRATION OF THE TUBE
PERMEATION TUBE METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
20
18
16
14
12
10
8
6
4
2 1
100ml
ppm
GAS To pump
DEHUMIDIFIER
(White)
105SD AMMONIA
Substance Interference Coexistence
Amines Similar stain is produced. Higher readings are given.
(NOTE)
When the concentration is below 1 ppm, 2 to 5 pump strokes can be used to detemine the lower concentration.
Following formula is available for actual concentration.
Actual concentration=Reading value× 1
Number of strokes
The KITAGAWA Gas Detector Tube System is a complete sampling and analysis system for determining hazardous gas and vapour concentrations quickly and easily. The KITAGAWA Gas Detector Tube System is comprised of an Air Sampling Pump and Precision Gas Detector Tubes.
The KITAGAWA air sampling pump reproducibly draws 100cc of sample air through the detector tube. All detector tubes are calibrated for a single pump stroke; there is no need for multiple volumes or stroke counters.
Each detector tube is formulated with high purity reagents that absorb and react with the target gas or vapour being measured. A colorimetric stain is created that is proportional in length to the concentration. For most tubes, the concentration is read directly off the measurement scale on each tube.
How To Use The KITAGAWA Gas Detector Tube System
Detector tubes are available to measure more than 200 different kinds of gases and vapours. For many gases and vapours, there are several concentration ranges available giving you the flexibility to measure both low and high concentrations without sacrificing precision.
How To Use The KITAGAWA Gas Detector Tube System
Detector tubes are available to measure more than 200 different kinds of gases and vapours. For many gases and vapours, there are several concentration ranges available giving you the flexibility to measure both low and high concentrations without sacrificing precision.
For specific information, please refer to the list of detector tubes.