产品详情
  • 产品名称:甲醛HCHO检测管NO.91PL

  • 产品型号:NO.91PL
  • 产品厂商:其它品牌
  • 产品文档:
你添加了1件商品 查看购物车
简单介绍:
甲醛HCHO检测管NO.91PL ,甲醛检测管NO.91PL,GASTEC甲醛HCHO检测管NO.91PL气体检测管,GASTEC甲醛HCHO检测管NO.91PL电泵用气体检测管检测范围0.01-0.80ppm。
详情介绍:
甲醛HCHO检测管NO.91PL
甲醛HCHO检测管
NO.91PL 甲醛检测管NO.91PLGASTEC甲醛HCHO检测管NO.91PL气体检测管GASTEC甲醛HCHO检测管NO.91PL电泵用气体检测管检测范围0.01-0.80ppm。
GASTEC为了满足用户在气体检测方面的需要,而开发出了电子自动采气泵。用传统的电子吸气泵几乎不能进行有效的采气,因为检测管里的填充试剂密度很高。而GASTEC开发的GSP电泵系列能提供强劲吸力,能满足特定的检测管的取样要求。电泵和检测管组成的新的检测系统能适用于作业环境的检测。

Formaldehyde

NO.91PL
HCHO 
甲醛HCHO检测管NO.91PL
Specifications
Calibration Range(0.01)-0.20 ppm
Measuring range0.01-0.80 ppm
Sampling (flow) Rate200mL/min
Sampling Volume6000mL, 2000mL
Sampling Time30min, 10min.
Detecting Limit0.005 ppm
Colour ChangeLight yellow → peach (pale pinkish) colour
Corrections for temperature/humidityTemperature
Shelf life1 year (in the refrigerator)
AccuracyCV=5-10%
(CV: Coefficient of variation = σ: standard deviation / mean value x 100)
Possibly coexisting substances and their interferences
Substance(s)InterferenceColour change by itself
Ammoniamax. 2 ppmNoNo
Nitrogen dioxidemax. 1 ppmNoNo
Acetaldehyde "+" error
(higher reading)
produces pinkish discolouration
Acetonemax. 0.3 ppmNoProduces pinkish stain around tube opening (inlet)
Ethanol NoNo
Ethyl acetate NoNo
p-dichlorobenzene NoNo
Toluene NoNo

Scrubber 1 (black):removes acetone

Scrubber 2 (purple):
Ammonia and nitrogen dioxide are removed. The presence of ammonia causes the purple scrubber 2 to turn yellow.

The presence of nitrogen dioxide causes scrubber 2 to turn a darker shade of purple
Calibration gas generationDiffusion tube method (analysis: by absorptiometry, acetylacetone method)
Miscellaneous
Electric sampling pump
This detector tube is suited for measuring formaldehyde in indoor environments. The GSP-200 / GSP-300FT series automatic gas samplers (or comparable sampling devices/pumps) should always be used for optimal performance.
A deviation of +/-10% in the sampling flowrate (i.e. amount of air flowing through the tube at a given moment) or the sampling amount (total sampled amount of target substance) causes an average margin of error of approx. +/-10% (maximum +/-20%) in the indicated value.
Formaldehyde indoor concentration threshold: 0.08 ppm (mean value during 30 minutes)
TLV-TWATLV-STELExplosive range
-- -- C0.3ppm7.0×73%